The MiniTest 3100 is no longer available and has been replaced by successor model MiniTest 4500/2500.
The all-rounder: the MiniTest 3100 can be combined with a variety of probes to make it suitable for use in many different areas. Measurement is non-destructive and covers the following ranges:
0 – 50mm for non-magnetic coatings on steel
0 – 100mm for insulating coatings on non-ferrous metals
10 – 200µm for copper coatings on insulating base materials
0 – 1600µm for powder coatings on steel and non-ferrous metals
Product infomation
Flexible application using the MiniTest 3100
The MiniTest 3100 system has numerous probes suited to a variety of coating thickness measurement tasks, as well as different calibration methods and analysis options:
Specialised probes can be supplied,g. completely sealed probes for use in wet areas, and heat-resistant probes for temperatures up to 250°C and 350°C
A special calibration method allows for countless different applications; calibration using unidentified coatings (CTC method) enables calibration on objects with inaccessible uncoated areas.
The calibration curve recorded in the factory is stored in the probe connector making it easy to interchange probes on site.
All probes are equipped with durable poles to ensure long life even for frequent use on abrasive substrates. The pole surfaces are highly polished.